YDB-03-BA-e-2114
97
when re-immersing into the liquid, no additional bubbles adhere; it is better to use
pincers or alike to place the sample directly on the sifting bowl.
Wait for stability display [ ], take over the weight value „sinker in test liquid“ using
the
button. The weighing balance will first determine then display the
fluid’s density.
When an optional printer is connected, the displayed value can be edited using
the
button.
Printout example (KERN YKB-01N):
d: 2.0000 g/cm
3
Return to weighing mode
Press the
ON/OFF
key
or use the
MENU
button to start a new measuring cycle.
If at the density determination errors have appeared, „d-----„ is displayed.
To avoid corrosion, don't leave the density set immersed in liquid for a long time.