5.9 KERN ADB / ADJ Series
The software of ADB / ADJ scale series does not have the density determination
function.
However, when combined with KERN Easy Touch software and <KERN SET-04
EasyTouch Density> application, it ensures easy, graphics-assisted determination of
solid body and liquid density (direct calculation and display of density).
Install KERN Easy Touch on your computer
(the pack of Basic ET license KERN SET-01 + KERN SET-04 application)
Using the appropriate RS-232C cable, connect the scale to the Com port of
your computer.
Trouble-free operation is ensured only when the appropriate interface cable by
KERN is used (optional).
Prepare the scale, see chapter 4.3.1
.
Install the density determination kit on your scale, see chapter 4.3.2
.
Remove the immersion basket and place a beaker filled with auxiliary liquid in
the plate center. It should be filled to ca. ¾ of its volume. It must not touch the
stand.
Hang the immersion basket again. It must not touch the beaker. Zero the
scale.
Determine the density, following the software instructions.
For further information / operating manual of KERN Easy Touch software see
the home page of KERN (