YDB-03-BA-e-2114
23
5. Put solid material into the lower sifting bowl.
For this remove the immersion basket out from the frame. Always ensure that,
when re-immersing into the liquid, no additional bubbles adhere; it is better to use
pincers or alike to place the sample directly on the sifting bowl. Make sure that the
sample is at least 1 cm immersed.
Figure 2
: Weighing in auxiliary liquid
6. Press the
UNIT
button. "WAIT" is displayed. The weighing scale first determines,
then displays the solid matter’s density.
7. If you connect an optional printer you can print the result.
8. Remove the sample. To carry out further measuring, press
UNIT
, and then start
with step 2.
To avoid corrosion, don't leave the density set immersed in liquid for a long time.