YDB-03-BA-e-2114
5. Place the solid body on the lower strainer pan.
To do it, remove the dipper basket from the stand. When submerging in the liquid
again, no extra air bubbles should be created. It is best to apply the sample using
tweezers or place it directly on the strainer pan. The sample must be submerged
at least 1 cm deep.
Fig. 2: Weighing in the auxiliary liquid
The sample weight in the auxiliary liquid will be displayed.
6. Wait until the stabilisation indicator is displayed (
), then press OK to take over
the weighed value. The density of the solid will be determined by the balance
taking into account aerodynamic displacement, and then displayed.
7. The result can be printed after an optional printer has been connected.
8. Remove the sample. To perform consecutive measurements, press
OK
and start
the procedure from step 3.
To prevent corrosion-related damage to the dipper basket, do not leave it
submerged in the liquid for an extended period of time.