YDB-03-BA-e-2114
49
Place sample of solid material in the lower sifting bowl, see fig. 2, chap. 5.1.2..
For this remove the immersion basket out from the frame. Always ensure that,
when re-immersing into the liquid, no additional bubbles adhere; it is better to use
pincers or alike to place the sample directly on the sifting bowl.
Make sure that the sample is at least 1 cm immersed.
Wait for stability display [ ], then take over the weight value „sample in aid liquid“
using the
button. The density of the sample is shown.
When an optional printer is connected, the displayed value can be edited using
the
button.
Printout example (KERN YKB-01N):
d: 2.0000 g/cm
3
Return to weighing mode
Press the
ON/OFF
key
or use the
MENU
button to start a new measuring cycle.
If at the density determination errors have appeared, „d-----„ is displayed.
To avoid corrosion, don't leave the density set immersed in liquid for a long time.