52
YDB-03-BA-e-2114
Place sample of solid material in the upper sample dish, see fig. 1, chap. 5.1.2.
Wait for stability display [ ], then take over the weight value using the
button.
Wait until the display for determination of „sample in aid liquid“ appears. Remove
the sample and if required tare by using the
TARE
button.
Lay the sample into the lower sample dish and immerse it in the aid liquid trying to
avoid bubble formation.
Make sure that the sample is at least 1 cm immersed.
Wait for stability display [ ], take over the weight value using
button. The
density of the sample is shown.
When an optional printer is connected, the displayed value can be edited using
the
button.
Printout example (KERN YKB-01N):
07-01-14 09:35:17
d: 8.0700 g/cm
3
To avoid corrosion, don't leave the density set immersed in liquid for a long time.