![KERN AEJ 120-4M Operation Manual Download Page 79](http://html1.mh-extra.com/html/kern/aej-120-4m/aej-120-4m_operation-manual_1957154079.webp)
AES/AEJ-BA-e-0912
79
14.4.5 Output weighing data in operating mode
S 1 = 1 word
S1
Code
Significance
L
4CH
Goods to be weighed below tolerance limit
G
47H
Goods to be weighed within tolerance range
H 48H
Goods to be weighed above tolerance limit
1- or 2 end points
1 31H
Limit 1
2 32H
Limit 2
3 33H
Limit 3
4 34H
Limit 4
5 35H
Limit 5
3- or 4 end points
T
54H
Total (add-up)
U
55H
Average piece weight (piece counting /
weight value (weighing)
(SP)
20H
No rating
d
64H
Gross
File Type
14.4.6 Data status
S 2 = 1 word
S 2
Code
Significance
S
53 H
Data stabilized *
U
55 H
Data not stabilized (fluctuating) *
E
45 H
Data error, all data apart from S 2 not allowed.
Balance indicating error (o-Err, u-Err)
sp
20 H
No special status
14.4.7 Interval data output
When an interval output is started or stopped, this will result in the output of a
header or footer line.
Header
consisting of 15 characters (CR=0DH, LF=0AH)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
- - - - - - - - - - - - - - -
Footers
Two line feeds are inserted.