![KERN AEJ 120-4M Operation Manual Download Page 51](http://html1.mh-extra.com/html/kern/aej-120-4m/aej-120-4m_operation-manual_1957154051.webp)
AES/AEJ-BA-e-0912
51
11.6 Adding of displayed values
This function can be used to automatically add any number of single weighings to a
total, such as single weighings of a batch.
The add-up function is available for the following functions:
•
Weighing
•
Parts
counting
•
Percent
determination
•
Weighing with freely programmable weighing unit
Activate function
Ö
In weighing mode keep the
F
-key pressed down
until
[
FWnm
]
appears on the display. Release but-
ton. The first menu item showing the current set-
ting will be displayed.
Fonm
Ö
Press the
F
-key and menu item
[
2
SEL
]
will
appear.
Ö
Select the desired setting by pressing the
TARE
-
key
[1]
: Add-up
[3]
: Tolerance weighing /Add-up
Ö
Press the
F
-key again
Ö
Select the desired setting by pressing the
TARE
-
key
[ 1 ]
: Add-up without AUTO-TARA
[ 2 ]
: Adding with AUTO-TARA
Ö
Confirm by pressing the
S
-key.
(
*
) indicates that weighing balance is in add-up
mode.