![KERN AEJ 120-4M Operation Manual Download Page 19](http://html1.mh-extra.com/html/kern/aej-120-4m/aej-120-4m_operation-manual_1957154019.webp)
AES/AEJ-BA-e-0912
19
8.5 Display speed
To change the display speed of the weighing balance without invoking the menu,
press the
S
-key. However, this only will limit your options to the three settings
nor-
mal
Æ
slow
Æ
fast
.
Within the menu it is possible to fine-tune the weighing balance for certain ambient
conditions and measuring purposes, see chpt. 10.1.1, menu items [
S.RE.
*] and
[
6.EnU.
*].
Ö
Press the
S
-key, current setting appears.
Ö
Repeated pressing of the
S
-key allows you to se-
lect any of the following settings
normal
Æ
slow
Æ
fast
Ö
After selecting, wait until the weighing balance
has returned automatically to weighing mode.
FAST
Ú
NORNAL
These settings correspond to the settings in menu item [
S.RE.
*] and [
6.EnU.
*].
Corresponding settings
in menu
S
-key
Adjustment Description
6.EnU
S.RE.
FAST
FAST
Weighing balance sensitive and
fast
0 1
nORNAL
NORMAL
Weighing balance working at
medium speed
0 3
SLQU
SLOW
Weighing balance insensitive and
slow
1 3
8.6 Data output
To print out or transfer weighing and GLP data to a computer you will have to con-
nect a printer or computer. The weighing balance’s factory settings define that data
output happens after successful rest position control was carried out followed by
pressing the
-key. For additional settings see chpt. 10.1.5.
During a display
weighing data are calculated via interface.