![KERN AEJ 120-4M Operation Manual Download Page 37](http://html1.mh-extra.com/html/kern/aej-120-4m/aej-120-4m_operation-manual_1957154037.webp)
AES/AEJ-BA-e-0912
37
11.1 Weighing / Weighing unit change
Ö
Place goods to be weighed on balance
Ö
Wait until the stability display appears [
|
]
Ö
Read weighing result.
Ö
Press repeatedly the button
F
to switch-over dis-
play into net/gross
Net weight unity A
↔
gross weight unity A
see chapter 8.4, or in other activated weighing
unity (menu setting
[
C3.W.B
]
)
100.0000
g
100.0000
g
B/G
500.000
mT
11.2 Parts counting
During piece counting parts can either be counted into a container or out of a con-
tainer. To count a greater number of parts the average weight per part has to be
determined with a small quantity (reference quantity).
The larger the reference quantity, the higher the counting exactness.
High reference must be selected for small parts or parts with considerably different
sizes.
The process has four steps:
•
Tare the weighing container
•
Determine the reference unit
•
Original weighing of reference weight
•
Count the items
Activate function
Ö
In weighing mode keep the
F
-key pressed down
until
[
FWnm
]
appears on the display. Release but-
ton. The first menu item showing the current set-
ting will be displayed.
Fonm
Ö
Use the
TARE
-key to select “piece counting” and
confirm by pressing the
S
-key.
The balance is situated in the piece counting
mode.
0
PCS