![KERN AEJ 120-4M Operation Manual Download Page 77](http://html1.mh-extra.com/html/kern/aej-120-4m/aej-120-4m_operation-manual_1957154077.webp)
AES/AEJ-BA-e-0912
77
14.4 Data output
14.4.1 Format for data transmission
You may customise the format of data transmission according to your requirements
(7-digit data format / extended 7-digit data format). See menu overview chpt. 10.2.
Setting ex-works:
7-digit data format, consisting of 16 characters, including terminator; CR=0DH,
LF=0AH (CR=weighing balance reverse / LF=vertical spacing). A parity bit can
be appended. The oblique “/” is printed just before the last digit.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
P1
D1 D2 D3 D4 D5 D6 D7 D8 D9 U1 U2 S1 S2 CR LF
14.4.2 Signs
P 1 = 1 word
P 1
Code
Significance
+
2 B H
Data is 0 or positive
-
2 D H
Data is negative
14.4.3 Numeric data
D 1 to D 8
8 words with 7-digit format
D 1 bis D 8
Code Significance
0 - 9
30 H – 39 H
Data 0 to 9
2 EH
Decimal point, position not fixed
Sp
20 H
Space character, leading cero suppressed