![KERN AEJ 120-4M Operation Manual Download Page 65](http://html1.mh-extra.com/html/kern/aej-120-4m/aej-120-4m_operation-manual_1957154065.webp)
AES/AEJ-BA-e-0912
65
•
Numeric entering of 2 limits
Example:
Reference weight
Lower limit
Upper limit
Weight value
100.0000 g
97.0000 g
105.0000 g
Input
100.0000 g
-3.0000 g
5.0000 g
1. How to enable the tolerance weighing function
[
2.SEL.2
]
or
[
2.SEL.3
],
see chpt. 11.7.3.
2. For settings in menu see chpt. 11.7.3.
Setting for 2 limiting values:
Setting for differential value:
3. How to save settings / how to exit the menu
Ö
Press the
S
-key
Weighing balance is in tolerance mode, the trian-
gular tolerance tags are faded in.
0.0000
g
4. How to set reference weight and limits:
Ö
Press the
S
-key for about 4 seconds until
[
r. SET.
]
appears and then let go. The most re-
cently saved reference weight will be displayed
(flashing).
z0000
g
M