11.3.2 Numeric entering of the reference weight
Activate function
Ö
See chpt. 11.3.1
0
%
How to determine a reference weight (100 %)
Ö
If required, place a weighing container on the
weighing pan
Ö
Press the
F
-key for about 4 seconds until
[
P. SEt.
]
appears and then let go.
The most recently saved reference weight will be
displayed (flashing).
P. SET.
.
O 0000
g
M
Ö
Enter reference weight (=100 %)
Each time the
TARE
-key is pressed, numbers will run
through from 0-9 and the decimal point.
⇒
0
⇒
0 1
⇒
⇒
⇒
≈
2 9
Select the number to be changed using
F
key (the
active position flashes):
Ö
Save the reference weight by pressing the
S
-key
or
cancel the entry by pressing the
-key.
000z0000
g
012z0000
g
Percent determination
Ö
Place goods to be weighed on balance.
The weight of the sample is displayed in percent-
age in terms of the reference weight.
8u37
%
Display change
Pressing the F-key repeatedly is a way to change the
display value, for instance to [g] or [%].