When
the
-key is pressed during the adjusting procedure, [stop] will
be displayed and adjustment interrupted. The balance returns to weighing
mode.
•
You can cancel the adjusting procedure by pressing any key.
•
The following error messages may be displayed during adjust-
ment/adjustment test.
1-Err
Incorrect adjusting weight
2-Err
Divergence last external adjustment > 1%
3-Err
During the adjustment there was one weight on the weighing plate
4-Err
Divergence from last internal adjustment > 1%
A-Err
Internal adjustment automatics defective
Adjustment protocols can be issued by connecting an optional
printer, see chpt. 15 "printer operation".
[
] is displayed during printing.
9.5 Advice CAL/Auto-CAL (KERN PEJ only)
Automatic adjustment is started or required every 4 hours, or when a temperature
change of 5 ºC occurs, depending on the setting in the menu, see chpt. 10.1.1.
Advice CAL = menu setting [
E. ADC. 1
]
When “Advice CAL“ is set, the required adjustment will merely be requested and
manual adjustment must be carried out.
“
CAL
” will start flashing if adjustment is required (after 4 h/ ± 5ºC). “
CAL
" will keep
flashing until the adjustment is started.
For AEJ models this is carried out with an internal adjustment weight (chpt. 9.1), for
AES models with an external adjustment weight (chpt. 9.3).
Advice CAL = menu setting [
E. ADC. 2
]
Automatic adjustment is carried out when “Auto-CAL“ is set and an adjustment due
(after 4 h / ± 5ºC).
“CAL” will start flashing 10 minutes before the start of automatic adjustment with
internal adjusting weight. Complete current weighing in processes within this time
frame. Automatic adjustment will not be started until the weighing balance is
unloaded and a successful stability display has appeared.
It is only possible to enable the Auto-Cal function in menu setting
[
x CA. 1
], see chpt. 10.1.2.