BOP-1K 031912
B-17
B.61 [SOURce:]LIST:DWELl? QUERY
LIST:DWEL?
Syntax:
Short Form: LIST:DWEL?
Long Form: LIST:DWELl?
Return Value: <value>
Description:
Identifies the dwell times entered for the list.
Starting at location established by LIST:QUERy,
returns comma-separated list of up to 16 values indicating the dwell time parameters entered. i.e., the
contents of LIST:DWEL locations of Table B-2. Related Commands: LIST: DWEL, LIST:QUERy. (See
example, Figure B-5.)
B.62 [SOURce:]LIST:DWELl:POINts? QUERY
LIST:DWEL:POIN?
Syntax:
Short Form: LIST:DWEL:POIN?
Long Form: LIST:DWELl:POINts?
Return Value: <value> (0 to 5900)
Description:
Identifies the number of locations for which time values have been entered and the next loca-
tion to be filled by a LIST:DWEL command.
If LIST:DWEL:POIN? returns 6, dwell times have been
entered for locations 0 through 5 and location 6 is the next to be filled by a LIST:DWEL command.
LIST:DWEL, LIST:DWEL:POIN. (See example, Figure B-5.)
B.63 [SOURce:]LIST:QUERy COMMAND
LIST:QUER
Syntax:
Short Form: LIST:QUER <int_value>
Long Form: LIST:QUERy <int_value>
int_value = <n> where n = first location to be queried.
Description:
Determines first location to be queried by LIST:VOLT?, LIST:CURR? or LIST:DWEL? queries.
Related Commands: LIST:VOLT?, LIST:CURR?, LIST:DWEL?. LIST:QUER?. (See example, Figure B-5.)
B.64 [SOURce:]LIST:QUERy? QUERY
LIST:QUER?
Syntax:
Short Form: LIST:QUER?
Long Form: LIST:QUERy?
Return Value: <int_value>
Description:
Identifies first location to be queried by LIST:VOLT?, LIST:CURR?, LIST:DWEL? queries.
Related Commands: LIST:QUER, LIST:CURR?, LIST:DWEL?. LIST:QUER?. (See example, Figure B-5.)
B.65 [SOURce:]LIST:REPeat COMMAND
LIST:REP
Syntax:
Short Form: LIST:REP <int_value1>,<int_value2>,<array>
Long Form: LIST:REPeat <int_value1>,<int_value2>,<array>
where
<int_value1> is start point to be copied
<int_value2> is end point to be copied
<array> is a series of comma separated values representing the new
programmed current (Amps) for current list, or programmed voltage (Volts) for voltage list.
Description:
Copies points from existing list using the same dwell times, but different output settings.
Cop-
ies the contents of the dwell entries in the list from the start point to the end point and adds them to the
end of the list.
The setting value for each copied point is defined by the value in the array. This process
is repeated until either a) an invalid programmed current or voltage is found, b) the all entries in the
array have been copied or c) the list is full.
This command must be preceded by LIST:CURR or LIST:VOLT command and the starting and end
point values must be valid, otherwise a -221 “Settings conflict” error results. The entries on the dwell
list must match the values in the current or voltage list, otherwise a -226 “Not Same Length” error
results. (See Figures B-6 and B-7.)
B.66 [SOURce:]LIST:RESolution? QUERY
LIST:RES?
Syntax:
Short Form: LIST:RES?
Long Form: LIST:RESolution?
Return Value: <value1>,<value2>,<value3>
(where <value1> and <value2> = clock period, <value3> = number of points to be used)
Description:
Identifies whether resolution is set to minimum (3933) or maximum (5900) and indicates the
clock period.
The resolution (3933 or 5900) determines the maximum number of points that will be
used for the waveform. <value1> and <value2> indicate the boundaries of acceptable dwell times:
e.g., if LIST:RES? returns either 0.000093,0.034000,3933 or 0.034000,0.000093,5900 dwell times
faster than 93
µ
S
and slower than 34 mS will not be accepted.
Summary of Contents for BOP-MG
Page 2: ......
Page 10: ......
Page 20: ...x BOP 1K 031912 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Page 37: ...BOP HIPWR 031912 1 17 FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Page 53: ...BOP HIPWR 031912 2 15 FIGURE 2 4 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Page 54: ...2 16 BOP HIPWR 031912 FIGURE 2 5 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Page 64: ......
Page 132: ......
Page 192: ......