38
SUPERtrol-I LE Flow Computer
Press Enter key to test the pulse input.
Use the Up/Down arrow keys to select the appropriate
trigger level.
Use the Up/Down arrow keys to select the appropriate
frequency range.
To check Pulse input accuracy: Use TB1-4 as reference
ground, input a frequency F1 on TB1-2, frequency F2 on
TB1-3. The display should show frequency being input.
Use a frequency counter to verify input.
Press Menu key to return to Pulse input test top-level
menu.
Press Enter to test the analog output.
To simulate analog output: Connect an ammeter to (+)
TB1-15, (-) TB1-16. Press the key under the desired
current setting to move the asterisk (*). The unit should
output the selected current.
Press Menu key to return to Analog out test top-level
menu.
Press Enter to test the excitation output.
To test the excitation output: Connect a voltmeter to
(+) TB1-1, (-) TB1-4. Press the key under the desired
voltage setting to move the asterisk (*). The unit should
output the selected voltage.
Press Menu key to return to Excitation out test top-level
menu.
8.2.8
Pulse input test
Sub-menu Group
8.2.10
Excitation out test
Sub-menu Group
2.5V
10mV
100mV
40Hz
3KHz
20kHz
8.2.9
Analog out test
Sub-menu Group
Pulse input test
Pulse input test
Trigger level 2.5V
Pulse input test
count speed 3kHz
Pulse input test
F1: 0 F2: 0
Pulse input test
Analog out test
Analog out test
*0 4 10 15 20 mA
Analog out test
Excitation out test
Excitation out test
*5v 12v 24v
Excitation out test
Notes
Display
Sub-menus
STOP
START
STOP
START
STOP
START
STOP
START