To set a custom default system speed mode, the FSE can add the following block to the top of the
icconfig_<QMO>.ini
file (example):
[SYSTEM SPEED MODE]
MODE=CUSTOM
SOURCE_SETTLING=3
NPLC=0.01
ANALOG_FILTER=1
DELAY_FACTOR=0
If you have problems with correlation of data using the system speed settings from the factory, your
FSE can add the following line to your
icconfig_<QMO>.ini
file to force the system to use the
older error mode:
TSP_ERROR_MODE OFF
(in KTE version 5.6.5 or later).
NPLC default differences
There are some instances where the number of power line cycle (NPLC) default values may be
different than what is specified in the
icconfig_<QMO>.ini
file or
KI_INTGPLC
setmode modifier:
The
meas
X
command always measures with an NPLC of
0.01
.
The default NPLC setting for the
intg
X
and
sintg
X
commands is
0.1
NPLC, except when
using the 10 nA and lower ranges on a 2636B (default NPLC of
1
).
The NPLC setting for measurements when running diagnostics is always
1
.
Decommissioning an S535 test system
The S535 Wafer Acceptance Test System does not contain any intentionally released substances,
but may contain substances that are potentially hazardous to the environment if not properly recycled.
For example, systems produced before July 22, 2017 and shipped into the European Union may
contain lead (Pb) as a part of the solder to connect electronic components and system interconnects.
A list of poss
ible hazardous substances is in European Union Directive “Restricting the use of
hazardous substances in electrical and electronic equipment” (RoHS) Directive 2011/65/EU or later.
This list includes:
Lead (Pb)
Mercury (Hg)
Cadmium (Cd)
Hexavalent chromium (Cr6+)
Polybrominated biphenyls (PBB)
Polybrominated diphenyl ether (PBDE)
Bis (2-ethylhexyl) phthalate (DEHP)