Appendix A. Specifications
Kaye ValProbe RT User Manual
138
A.3.8
Housing Material
Pharma grade PEEK (Polyetheretherketone) material housing and conforming to safety
UL 94-V-0 flammability rating.
A.3.9
Logger Base Dimensions
Height 48mm (1.9”) / Diameter 36 mm (1.4”).
A.3.10
Ingress Protection
IP68 compliant.
A.3.11
Insulating Canister X2545
Size 149mm (5.87”) x 45mm (1.75”) (Length x diameter), weight 330grams (0.73 lb).
A.3.12
Battery
Field replaceable 3.6VDC ½ AA Lithium Thionyl Chloride primary battery.
A.3.13
Shelf Life
> ten years.
A.3.14
Battery Life
Battery life is affected by several factors, including sample rate, transmission rate,
temperature, logger type and hours of use. Individual battery capacity can vary from one
battery to another by up to 10%.
A.4 Kaye ValProbe RT Temperature Logger
A.4.1
Measurement Accuracy
±0.1°C from -85 to 140°C.
±0.25°C from 140 to 400°C for bendable models.
A.4.2
Body Environmental Ratings:
Temperature: -85 to 140°C.
Temperature: -85 to 360°C with canister model.
Humidity: 0 to 100%, condensing.
Pressure: 0 to 5 bar absolute.
A.4.3
Number of Input Channels
1, 2 and 5 channels (5ch logger sampling rate is min two secs).
A.4.4
Sensor Types and Measurement Range
Rigid, flexible, bendable and surface mount: -85 to 140°C.
Summary of Contents for ValProbe RT
Page 1: ...Validation Kaye ValProbe RT User Manual MXXX EN Rev 1 0 May 2020 ...
Page 2: ... No content intended for this page ...
Page 4: ... No content intended for this page ...
Page 12: ......
Page 163: ...Kaye ValProbe RT User Manual 150 ...
Page 166: ... No content intended for this page ...