Chapter 2:
Kaye ValProbe RT Hardware
Kaye ValProbe RT User Manual
21
Base Station IRTD / Bath Connections
At the rear of the Base Station there are connections for the IRTD and Kaye
References which can be used for Sensor Verification. Using the supplied
cables connect the IRTD and bath to the appropriately marked connectors.
The provides cables and connections are backward compatible to all Kaye
baths, dry block and IRTD’s. The ValProbe RT software can provide fully
automated sensor verification of your ValProbe RT loggers (See Chapter 10
for details on Sensor Verification).
Figure 31: IRTD / Kaye Bath Connections
Base Station Battery Backup
Kaye ValProbe RT base station has battery backup for up to 60 minutes. The
Base Station is powered by two 3.7V AA size rechargeable Lithium batteries.
The Base Station has a built-in charger, so batteries will be charged
automatically when there is a main supply available to the base station. To
replace the batteries, first shut off the base station using ON-OFF switch. Then
open the sliding panel on the back of the base station and replace the
batteries in the compartment displayed in Figure 32 below. Do not power on
the Base Station until you have properly replaced the compartment cover.
Figure 32: ValProbe RT Base Station Battery Compartment
CAUTION!
ONLY Use Kaye #200-166 3.7 V lithium secondary
(rechargeable) batteries. Using any other batteries renders
the stated performance void as Kaye #200-166 3.7V batteries
are the only brand tested and accepted.
Insertion of non-rechargeable battery may cause over-
heating and explosion.
Summary of Contents for ValProbe RT
Page 1: ...Validation Kaye ValProbe RT User Manual MXXX EN Rev 1 0 May 2020 ...
Page 2: ... No content intended for this page ...
Page 4: ... No content intended for this page ...
Page 12: ......
Page 163: ...Kaye ValProbe RT User Manual 150 ...
Page 166: ... No content intended for this page ...