Chapter 2:
Kaye ValProbe RT Hardware
Kaye ValProbe RT User Manual
18
Important:
I
f the LED does not blink, remove battery and wait (1-2 minutes)
until all the residual power is released. Then use a magnet over
the Logger without the battery. Reinstall the battery and wait for
the LED to blink three times. If LED keeps blinking, remove the
battery again and then reinstall the battery. If the Logger blinks
more than three times, the Logger may have an issue with its
hardware.
•
Reinsert the Logger cover and hand tighten until the cup O-ring is no longer
visible. Clean the exterior of the Logger with a soft cloth to remove any excess
lubricant.
Reset the Battery Odometer
When you change the battery in a Logger, you must also reset the battery
odometer to 100% for the program to calculate the life of the new battery.
Reset Battery Life:
•
Start the ValProbe RT software.
•
Navigate to the Hardware Discover tile, select a Base Station and then select
the
Connect
button.
Figure 25: Connect Button
•
Wake up all Loggers that have had battery replacements using the Wakeup
device or the Base Station magnet.
•
From the Logger Configuration screen, select the Loggers and then
Battery
Reset
.
Figure 26: Select Loggers