9.3.2
On the unit 230 V version
▪ Requirements for the power cable, the device-side and mains-side plugs:
– 230 V rated voltage
– 16 A rated current
– Length 3 m
▪ Power cable H05VV-F3 1.50 (3 x 1.5 mm²), unshielded
▪ Device-side plug in accordance with IEC 60320 C 19.
▪ Mains-side plug, Schuko plug, straight, in accordance with CEE (7) VII
▪ Important, protective conduct or connection!
▪ The components that are used must satisfy these requirements as well as coun‐
try-specific requirements.
9.3.3
On the unit 100-127 V version
▪ securely connect the power line to the interfering voltage transformer or isolating
transformer
▪ Mains-side plug, rated voltage with 25 A or 20 A for Japan.
▪ Important, protective conduct or connection!
▪ Mains connecting plug in correspondence with country-specific requirements
▪ Mains-side plugs for USA only 125V/30AMP NEMA 5-30P, UL-listed, CSA-listed
9.4
Transportation and storage
9.4.1
In the dental practice
CAUTION
Damage from transport
Property damage
▶ Only pull the KEY Laser 3+ / III upgraded on the grip along the device when
transporting it.
▶ When transporting it over a threshold, only pull in the direction of the arm and
slightly lift the device.
9.4.2
Outside of the dental practice
The KEY Lasers 3+ / III upgraded comes in cardboard packaging.
The instructions for transportation and storage printed on the outside must be ob‐
served.
Transport upright: The arrow indicates the top.
Protect from impact.
Protect from moisture.
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
9 Assembly instructions | 9.4 Transportation and storage