In addition, the triangular laser warning sign and instructions sign are affixed where
the laser hose is connected.
Laser aperture
at
distal end
3.5
Technical data and requirements
Setup plan No.
111 052 94 003
Electrical lead
230 V Schuko socket
Mains power input line
3 m long
Input voltage
230 V~
By installing the "External voltage kit" or
an isolating transformer, the input volt‐
age is expandable to
100 V~/110 V~/120 V~/127 V~
Frequency
50/60 Hz
Power consumption
40 - 2200 W and 100 - 2500 VA
Customer fuse protection 220-240 V
B16 or C16- automatic
Customer fuse protection 100-127 V
25 A
Customer fuse protection 100-127 V
(Japan)
20 A
Classification
Device of protection class I
Device not protected against the pene‐
tration of water
Type B application part
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
3 Device description | 3.5 Technical data and requirements