Heat emission
0.4 – 8 KJ/h
Cooling
air-cooled with internal water circuit
Cooling water
Conductivity below 10 µS/cm (refer to
VDE 0510)
Amount of cooling water
approx. 2.5 l
Spray water supply
Reservoir integrated in the KEY Laser
3+ / III upgraded
Spray water quality
only distilled and demineralised water,
pH-neutral
Spray water level
100 - 400 ml
Spray air supply
integrated in the KEY Laser 3+ / III up‐
graded
Unit weight
78 kg
Connection to door contact
max. 5 V / 10 mA (load on the door con‐
tact)
Switching contact for external warning
lamp
max. 24 V / 1A
Packaging, transportation and storage
Gross weight
93 kg
Net weight
81 kg
Package dimensions
1100 x 600 x 900 mm
Transportation and storage conditions
temperature
min.: - 20 ℃, max.: + 55 ℃
Humidity
min.: 5%, max.: 95 %
Air pressure
min.: 700, max.: 1060 hPa
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
3 Device description | 3.5 Technical data and requirements
29/76