9 Troubleshooting
Malfunction
Cause
Remedy
Camera image freezes
without the release
button or foot control
having been triggered.
Taking the camera out
again did not solve the
problem.
Electrical or electro-
magnetic interference
by other equipment.
▶
Restart the software.
Camera image freezes
without the release
button or foot control
having been triggered.
The monitor turns itself
off.
Electrical or electro-
magnetic interference
by other equipment.
▶
Restart the treatment unit and the CONEXIO
PC.
An acoustic signal is is-
sued every second.
Leaking water switch
recognises leaking wa-
ter.
▶
Remove water from the unit body. If neces-
sary, have a technician fixed the leak.
Malfunction
Cause
Remedy
Display shows: ID 33
No CAN node and/or in-
ternal communication is
faulty.
▶
Call a service technician.
Display shows: ID 64
Water is shut off.
▶
Turn water on.
Water works leaks
strongly.
Water works malfunc-
tion
▶
Call a technician.
Display shows: ID 65
Safety switch of bowl
suction has been
reached.
▶
Turn external suction on.
▶
Check and clean, if required, the bowl valve.
Display shows: ID 66
Amalgam separator
malfunction
▶
Remedy the malfunction.
See also:
2
Instructions for Use of amalgam separator
Display shows: ID 67
The oxygenal container
is empty.
▶
Refill the Oxygenal container.
See also:
2
Servicing instructions
Display shows: ID 68
Call for service
▶
Have a service performed.
▶
Call a technician.
Display shows: ID 69
Intensive germ reduc-
tion must be done.
▶
Carry out an intensive germ reduction.
See also:
2
Servicing instructions
117 / 126