4 Operation | 4.2 Adjusting the dental chair
Saving an automatic position with the foot control
①
Spray preselection/AP footswitch
②
Blown air/AP footswitch
③
Foot pedal
The chair positions can be saved on two footswitches; the standard setting is as
follows:
▪ "Spray default" footswitch: "LP" automatic position (last position)
▪ "Blown air" footswitch: "SP" automatic position (rinsing position)
▶
Hold down the foot pedal and foot-operated button "SP", and simultan-
eously press any button for an automatic position ("AP 0", "AP 1", "AP 2" or
"SP") on the dentist or assistant unit until you hear a beep.
ð
The automatic position is saved to the foot-operated button.
or
▶
Hold down the foot pedal and foot-operated button "LP", and simultan-
eously press any button for an automatic position ("AP 0", "AP 1", "AP 2" or
"SP") on the dentist or assistant unit until you hear a beep.
ð
The automatic position is saved to the foot-operated button.
4.2.5 Safety shut-off
To prevent collisions arising from the movement of the patient chair, safety
shutoff switches are installed to protect the patient and practice personnel from
injury and the treatment centre from damage.
NOTICE
Damage to the assistant element and dental chair.
Despite some safety shut-downs being present, certain positions of the
assistant element may collide with the dental chair.
▶
Keep the assistant element out of the motion range of the patient chair.
▶
Always monitor the chair movement.