Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
106 / 126
Note
In the testing of ME devices with several application parts, the parts must be
connected in succession. The measured results must be evaluated using the
threshold values. Application parts, which are not included in the measure-
ment, remain open.
Note
An additional measurement of the leakage current from type B applied parts
need only be carried out if this is specified by the manufacturer (see accom-
panying documents).
Note
A separate measurement is not usually required for type B applied parts. The
applied parts are connected to the casing (see diagram) and included in the
measurement of the leakage current of the casing, whereby the same per-
missible values are applicable.
7.2.4 Functional test
The following conditions must be met in all function tests:
▪ The basic function of the treatment centre must be guaranteed.
▪ The treatment centre must be fit for use.
▪ There must be no irregularities, noise or abrasion, etc., present.
The following list is for exemplary purposes and makes no claim of being com-
plete.
▪ Functional test of the safety circuits (see diagram below)
▪ Functioning of the master switch of the device
▪ Functioning of the displays
▪ Functional test of the holder switch of the dentist and assistant element
▪ Functional test of the 3F/MF handpiece – seating of the cannula
▪ Functional test of the operating light
▪ Functional test of the suction hoses
▪ Functional test of the foot control
▪ Function of the chair:
- Travel on all axes
- Testing of the limit switches
▪ Functional test ...