Instructions for use Primus 1058 Life
3 Product description | 3.11 Technical data
40 / 126
Air supply
WARNING
Non-compliance with national guidelines concerning the quality of the
dental air.
Infection hazard.
▶
Note and comply with the national guidelines concerning the quality of the
dental air - if any.
▶
Blow through the air line prior to commissioning.
Air inlet pressure
5.2 to 7 bar
Minimum air flow rate
min. 80 Nl/min
Pressure dew point
< -30 ºC (compressor with
dry air system)
Oil content
< 0.1 mg/m
3
(oil-free compressor)
Contamination
< 100 particles/m
3
at particle sizes
of 1 to 5 µm
Customer-site air filtration
50 microns
Air connection
Shut-off valve with brass cone com-
pression screw connection 3/8" to Ø
10 mm provided
Air connection above floor level
min. 50 mm, max. 105 mm with
valve opened
Suction
Pressure drop at the connecting point
①
Semi-dry and wet suction
②
Dry suction
Note
If the negative static pressure is > 180 mbar, the unit must be equipped with
the negative pressure regulating valve assembly kit.