Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
70 / 126
▶
Press the "Increase value" key to increase the speed.
ð
The intensity is shown on the display.
▶
Press the "Save" key to save the values. You can save after setting each
value, or after setting all values.
ð
Saving is acknowledged by a beep.
ð
This closes the "Settings" menu.
Setting the cooling level
Requirement
Settings menu for turbine is selected.
▶
Press the "Preselected spray" button.
or
T
M
▶
Press the "Preselected spray" footswitch. (with level switching only)
ð
The set values are saved each for the set memory level and the set dentist
level.
Key
Feature
No LED is on: No cooling
One LED is on: Spray air cooling status
Both LEDs are on: Spray cooling status
Changing motor settings in the menu
Note
Following instructions for use, service instructions and installation instructions
in the instrument packaging.
Note
The motor mode is equivalent to 2 minutes operating time and 5 minutes
pause.
This represents the possible maximum load of the motor (full load at max-
imum speed).
In practice, pulse loads lasting seconds or pause times lasting seconds or
minutes are realistic given that the maximum possible motor current is not
normally reached. This equates to the dentist's normal way of working.
▶
Remove the motor from the holder.
▶
Briefly press the "Preselect level" (S1) key to select the level.
▶
Press the "Preselect level" (S1) key for 4 seconds to change settings.
ð
The display switches to the motor settings menu.