Instructions for use Primus 1058 Life
9 Troubleshooting
116 / 126
Malfunction
Cause
Remedy
Beeps every
10 seconds.
The LED on the "In-
tensive germ reduc-
tion" key flashes
(green). (Assistant ele-
ment) MEMOspeed/
men displays an error.
The Oxygenal container
is empty.
▶
Refill the Oxygenal container.
See also:
2
Servicing instructions
The light of the operat-
ing light switches off
during treatment and
re-starts after approx.
5 seconds.
Voltage fluctuations in
the supply mains
▶
If the light switches on automatically, there is
no need for further action.
LED on "HYDROclean"
button (red) flashes.
Malfunction in the am-
algam separator.
▶
Call a service technician.
▶
Note the amalgam separator warning notice.
See:
Operating Instructions of the amalgam
separator
Emergency shut off of
the bowl valve (only
when external suction
is installed)
▶
Call a service technician.
ERGOcam does not
work.
PC is switched off.
▶
Turn on the computer.
USB cable too long.
▶
Make sure that the cable length does not ex-
ceed 10 m (2 x 5 m passive with repeater).
No data transmission to
the multimedia menu
of the unit.
No or faulty ethernet
connection between
dental unit and office
network.
▶
Notify network administrator.
Camera images shows
images only as black/
white images.
Electrical or electro-
magnetic interference
by other equipment.
▶
Restart the CONEXIO PC.
Camera image freezes
without the release
button or foot control
having been triggered.
Camera image fails to
return to live image
mode.
Electrical or electro-
magnetic interference
by other equipment.
▶
Replace the camera in the holder and then
take it out again.