Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
54 / 126
CAUTION
Pinching from the treatment chair.
The safety shutoff of the treatment chair is activated by lifting the respective
component. Depending on the patient's body weight and the leverage, more
force can be exerted on the object to be triggered than is necessary to trigger
the switching function.
▶
The treatment personnel must move outside of the chair's swivelling range
whenever the chair moves.
The safety cut-offs can be found at the following places on the treatment unit.
Safety shutoff for the standard patient chair
①
Patient unit pivoted over the patient
chair
②
Assistant element
③
Backrest
④
Bracket on the foot control
⑤
Kick plate
⑥
Seat
Item
No.
Safety switch-off ac-
tuated
LED on assistant
element
LED on dentist ele-
ment
①
Patient unit pivoted
over the patient chair
②
Assistant element
③
Backrest
④
Bracket on the foot
control
⑤
Base plate
⑥
Bench