REMOTE/AUTOMATED CONTROL 6-19
_______________________________________________________________________
Copyright © 2010 by Kaufman & Robinson, Inc., 1306 Blue Spruce Drive, Unit A, Fort Collins, CO 80524
Tel: 970-495-0187, Fax: 970-484-9350, Internet: www.ionsources.com
mand can be used to set or query the values for individual parameters or
for a string of parameters depending on the format of the command.
The general form of the program read/write command is:
Command: P<program#>:<parameter> (value)
The <parameter> and (value)
fi
elds are optional. The command
can be turned into a query by substituting ? for the <program#>,
<parameter>, and (value)
fi
elds. The interpretation of the query is
dependent on where the ? appears as outlined below.
●
Set or query the active program number
P<program#><cr><lf> - Set active program to <program#>
P<?><cr><lf> - Get active program number.
Acceptable program numbers would be 1 to 4.
Any parameters following the ? character for this command will be
ignored.
Any parameters following <program#> without the <:> as a delimit-
ing character before the <parameter>
fi
eld will be ignored.
Example: P4<cr><lf> - Set active program number to 4
Example: P?<cr><lf> - Get active program number
●
Set or query the values for individual program parameters.
P<program#>:<parameter> (value) – Set speci
fi
c <program#> <pa-
rameter> to (value)
P<program#>:<parameter> (?) – Get speci
fi
c <program#> <param-
eter> (value)
There are 7 <parameters> for setting an individual (value). Each
of these <parmeters>, which are three letters long, are listed below
along with the description and the number of signi
fi
cant digits and
decimal places shown as x: