4.3.2 Gas Only Mode

When gas only mode is selected, the Auto Controller functions as a gas 

controller only. Independent control of the each of the four gas channels 

is possible. The description below outlines the use with one gas channel 

even though similar operation could be obtained using all the gas 

channels. Use the Operating Mode Select button to select the gas only 

mode. Selecting the Gas Only mode releases control of the Filament and 

Discharge Power supplies so that they can be operated independently. 

There is no feedback to maintain the discharge parameters. However, 

at any time during the operation in gas only mode, the setpoints can be 

acquired for operation later using the Auto Gas mode or the Manual Gas 

mode.

Note that if there is no communication link to either the Filament Power 

Supply or the Discharge Power Supply, the Auto Controller will only 

function in the gas only mode, which must be selected to clear the HLP 

7 message from the display. There will be no communication link if the 

power is switched off to either of the power supplies or if any RS-485 

cable connection on the rear panel is not connected to Auto Controller or 

either of the power supplies.

There are two modes of operation for the discharge: Constant Current 

Mode or Constant Voltage Mode. The Constant Current Mode allows for 

small variations in the discharge voltage while maintaining a constant 

current and is recommended when operating at 120 V and below. 

The Constant Voltage Mode will allow for small variations in discharge 

current and maintains a constant voltage. Constant Voltage Mode is 

recommended for operation above 120 V.

4.3.2.1 Constant Current Mode

The operating condition selected for this demonstration is with a 2.5A, 

120 V, discharge, which is Constant Current Mode operation. These 

conditions are based on a vacuum pump speed of 800 liters per second. 

Operating parameters for other pump speeds and gases can be found in 

the Ion Source Manual.

  Note that the operating range of the ion source may be limited by the 

vacuum facility pump speed or other process that take place while the 

ion source is running.

• Select Gas 1 using the white Gas Channel Select button on the Auto 

Summary of Contents for eHF 3005A

Page 1: ...350 Internet www ionsources com Copyright 2011 by Kaufman Robinson Inc All rights reserved No part of this publication may be reproduced without written permission February 2011 VERSION 4 AUTOMATED CO...

Page 2: ...2 3 4 Connections Between Controller Components 3 2 3 5 Connect Controller to Ion Source 3 5 4 INITIAL OPERATION 4 1 4 1 Gas 4 1 4 2 Power On 4 1 4 3 Operating Mode 4 1 4 3 1 Auto Gas Mode 4 1 4 3 1 1...

Page 3: ...m Good on the DB 37 Connector 6 7 6 2 6 Description of the DB 37 pins 6 8 6 3 Interlock 6 11 6 4 RS 232 Serial Interface 6 12 6 4 1 Interface cable 6 11 6 4 2 Communication settings 6 11 6 4 3 Communi...

Page 4: ...Controller Inventory List 3 1 3 2 Input Power Requirements 3 2 3 3 Pin descriptions for the rear panel DB 15 female MFC output connectors 3 4 6 1 Pin descriptions for the female DB 25 connector on th...

Page 5: ...s CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com 1 SAFETY Caution High Voltage Only technically qualified personnel should install maintain and troubleshoot the equipment descr...

Page 6: ...eight and dimensions of each controller are outlined in the following table This section gives a general overview of the Auto Controller package with more detailed information in the later sections of...

Page 7: ...s flow controllers using cables that connect to the rear panel Four relay contacts are also available on the rear panel if additional positive shut off valves are desired for the gas feed system The A...

Page 8: ...ty of mass flow controllers How ever the cables may vary depending on the mass flow controller provided in the order The pin descriptions for the female DB 15 connectors on the back of the Auto Contro...

Page 9: ...S _______________________________________________________________________ Copyright 2010 by Kaufman Robinson Inc 1306 Blue Spruce Drive Unit A Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 I...

Page 10: ...e Unit A Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com GENERAL DESCRIPTION AND SPECIFICATIONS 2 5 MFC Cable Auto Controller Filament Discharge Operating Cable Ope...

Page 11: ...n addition to the shipping company to report the damage see Section 8 Retain packaging materials for any future shipments Table 3 1 is an inventory list for the controller package The inventory list i...

Page 12: ...nt Power Supply 115 V 1 50 60 Hz 208 230 V 1 50 60 Hz 15 A 7 5 A Discharge Power Supply 115 V 1 50 60 Hz 208 230 V 1 50 60 Hz 15 A 15 A The discharge power supply output will be limited to 1500 watts...

Page 13: ...the rear panel of the Discharge Power Supply and tighten the screw above the appropriate connector Attach the ground connections using the green striped Yellow wire with the three ring terminals as f...

Page 14: ...r is used that cable can be attached at this time to either the female DB 25 connector the female DB 37 connec tor or the female DB 9 connector These remote interfaces require knowl edge of the ion so...

Page 15: ...the Installation Inspection description in the ion source manual The ion source installation procedures cover installation of the ion source vacuum cables gas tubing inside the vacuum chamber as well...

Page 16: ..._______________________________________________ Copyright 2010 by Kaufman Robinson Inc 1306 Blue Spruce Drive Unit A Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com...

Page 17: ...RS 232 Remote iMII Analog Remote KRI Analog Remote INSTALLATION 3 7 _______________________________________________________________________ Copyright 2010 by Kaufman Robinson Inc 1306 Blue Spruce Dri...

Page 18: ...act Kaufman Robinson Inc The bottle regulator and mass flow controller s should be installed and gas lines evacuated as described in Section 2 4 3 of the Ion Source manual 4 2 Power On Switch on the f...

Page 19: ...moves to the next program etc Cycling around to the program that s currently running will continue operation without change Otherwise the parameters will be ramped to the newly selected program 4 3 1...

Page 20: ...current setpoints can be adjusted using the Setpoint Adjust knob on the front of the discharge controller These adjustments can be made when the Auto Controller is in Standby or Enable The setpoint b...

Page 21: ...discharge current when the auto controller is enabled also adjusts the emission current In this case the emission becomes a fixed ratio of the emission current to discharge current based on the emissi...

Page 22: ...ler is enabled or when it is in standby When the Auto Controller is in standby use the program select button to select the program number that will be overwritten Adjust the setpoints to the desired v...

Page 23: ...function in the gas only mode which must be selected to clear the HLP 7 message from the display There will be no communication link if the power is switched off to either of the power supplies or if...

Page 24: ...set to 2 5 amps as indicated on the Discharge Amps display Press the white Enable Standby button on the Auto Controller Press the white Enable Standby button on the Filament Power Sup ply Press the wh...

Page 25: ...2 5 A 150 V discharge which is Constant Voltage Mode operation The following conditions are based on a vacuum pump speed of 1600 liters per second Operating parameters for other pump speeds and gases...

Page 26: ...Manual Gas mode To do this first press the white Program Select Button repeatedly until the desired program number is selected as indicated by the numbered red LED s Next press the white Acquire Setp...

Page 27: ...7 and 20 will put the unit into standby If this option was requested with the initial order then a male DB 25 pin connector with two wire pigtails or a user specified connector will be supplied Supply...

Page 28: ...rge Power Limit The discharge power supply has a limit indicator LED This power limit LED in dicates possible overheating of the ion source components See the ion source manual for the power limit tha...

Page 29: ...art of the ion source for electrical operation is the anode and the filament cathode The ion source body is grounded and no isolator is re quired in the gas line connection to the ion source More comp...

Page 30: ...____________________________________________________________________ Copyright 2010 by Kaufman Robinson Inc 1306 Blue Spruce Drive Unit A Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Intern...

Page 31: ...enable the system using the rear port These programs can be acquired using the front panel as described in the operation sections 4 and 5 of this manual or they can be acquired using the rear panel re...

Page 32: ...esents full scale output from the MFC ref erenced to pin 10 2 Emission Current Actual 0 to 5V DC output where 0V DC represents 0A DC of emission current and 5V DC represents 7A of emission current ref...

Page 33: ...n 12 No Connection 13 No Connection 14 Anode Current Actual 0 to 5V DC output where 0V DC rep resents 0A and 5V DC represents 5A DC referenced to pin 10 15 Cathode Filament Current Actual 0 to 5V DC o...

Page 34: ...orts When using remote enable the programs can be acquired using the front panel as described in Operation sections 4 and 5 of this manual or the can be acquired remotely as described below in sec tio...

Page 35: ...int using Manual Gas Mode The remote gas setpoints are not available when using the manual gas mode instead the starting gas flows are established by preprogramming them from the front panel for each...

Page 36: ...n Reference for gas channel setpoints See section 6 2 3 1 below 22 Discharge Current Setpoint 0 5 volts corresponds with 0 to the maximum level of discharge current in amps for the Discharge Power Sup...

Page 37: ...maximum output in sccm for the channel 1 MFC Refer enced to pin 27 10 Channel 3 Gas Flow Output 0 5 volt corresponds with 0 to the maximum output in sccm for the channel 3 MFC Referenced to pin 27 20...

Page 38: ...ction gives a real time check of the operating param eters If the operating parameters are in range the two Beam Good pins are shorted together If the operating parameters are out of range then the be...

Page 39: ...troller being used Referenced to pin 7 or 8 5 Auxiliary 12 volts Referenced to pin 7 or 8 6 Filament Voltage Output 0 5 volts corresponds with 0 50 volts for the Filament Power Supply Referenced to p...

Page 40: ...h 0 to the maximum level of discharge voltage in volts for the Dis charge Power Supply that is being used 22 Discharge Current Setpoint 0 5 volts corresponds with 0 to the maximum level of discharge c...

Page 41: ...Loc The red Open Interlock indicators on the front panels of the Fila ment Power Supply and the Discharge Power Supply are also on when the Auto Controller interlock is open if there is an RS 485 com...

Page 42: ...appendix if another baud rate is required 6 4 3 Communication overview The acceptable command response behavior for the Auto Controller is described below The action reaction of the Auto Controller to...

Page 43: ...ibed below in the Verbose Command section All commands are expected to be terminated with a CR LF indi cated by cr lf in the description below 6 4 4 1 VRB Verbose Command The VRB command sets a flag f...

Page 44: ...n Once Enabled the RS 232 remote control locks out all front panel controls except for the ON OFF switch and Gas Select The Gas Select button is left active to allow monitoring of individual gas flow...

Page 45: ...emote control Example COM 1 cr lf Enable RS232 remote control Example COM cr lf Get present output state If COM is not Enabled then the response to commands will be Comm Inactive in both Terse and Ver...

Page 46: ...ault 6 4 4 5 TST Self Test Query The TST command performs a unit self test and report results A non zero return value will correspond to the associated front panel error message as outlined below If m...

Page 47: ...Standby Example OUT 1 cr lf Enable the unit Example OUT cr lf Get present state of output If the unit is in Verbose the response will be Standby or Enabled If the unit is in Terse the response will be...

Page 48: ...as or Gas Only If the unit is in Terse the response is 0 1 or 2 6 4 4 8 BEAM Beam Good Query The BEAM query Indicates if a stable operating point has been achieved Command BEAM Response is 1 if beam g...

Page 49: ...ere the appears as outlined below Set or query the active program number P program cr lf Set active program to program P cr lf Get active program number Acceptable program numbers would be 1 to 4 Any...

Page 50: ...will be truncated Example P4 GS1 10 cr lf Set program 4 channel 1 gas flow to 10 sccm Example P1 DSV cr lf Get program 1 Discharge voltage value Example response in Terse or Verbose mode 200 000 Set...

Page 51: ...mple P1 ALL GS1 GS2 GS3 GS4 DSV BEI cr lf When using the ALL command in Terse mode to get all the program values they will be returned as a comma delimited string of values in the same order as those...

Page 52: ...mmand Issuing the OUT 0 command will dis able the gas In using the ALL parameter to set targets the list of values must be comma delimited and appear in the order shown below Any value that is to rema...

Page 53: ...rameters as those described above for the P command except that there is no program number required and the character is not used R parameter Get the current running value of parameter Values are repo...

Page 54: ...RN flag Setting the LRN flag with COM in active is ignored Setting the LRN mode using the serial command temporarily overrides the LRN mode setting established in setup The value in setup is restored...

Page 55: ...ntroller The auto controller front panel display has an additional function of displaying help codes and error messages If it becomes necessary to display a help code or an error message the Auto Cont...

Page 56: ...ror Message Description Possible Causes And Corrective Action HLP 8 RS 232 Interface Serial Common Error High current flowing in the ground connection of the serial interface between the unit and the...

Page 57: ...signal to clear the code The unit remains in STANDBY until STANDBY ENABLE is pressed again or the remote enable signal is reap plied Anode has an insulating coating and needs to be cleaned Filament i...

Page 58: ...signal to clear the code The unit remains in STANDBY until STANDBY ENABLE is pressed again or the remote enable signal is reap plied Filament is broken and needs to be replaced In this case HLP 3 is d...

Page 59: ...ze during the start up se quence Press and release the STAND BY ENABLE button or remove the remote enable signal to clear the code The unit remains in STANDBY until STANDBY ENABLE is pressed again or...

Page 60: ...turned on Int LOC Open Interlock An open interlock prevents the Auto Controller from go ing to the ENABLE state and holds it in STANDBY This is a user defined fault in remote operation Review the inte...

Page 61: ...d off to prevent a failure Switch off the power until the display is dark and then switch the power back on to reset HLP 2 Thermal Failure Open or shorted thermistor detected Contact KRI HLP 3 Broken...

Page 62: ...the power supply turned on until the Hot message is cleared Make sure there is no blockage of the vents on the rear panel Ambient temperature is too high Reduce ambient temperature below 40 C Not Rdy...

Page 63: ...t Check connection of the RS 485 cables Make sure Auto Controller has input power and is turned on HLP 8 RS 232 Interface Serial Common Error High cur rent flowing in the ground connection of the seri...

Page 64: ...over ordinary wear and tear or expendable components KRI s obligation under this Limited Warranty is at KRI s option to repair or replace any defective equipment or parts of the equipment without char...

Page 65: ...epairs or replacement during Warranty or repairs there after please contact Kaufman Robinson Inc 1306 Blue Spruce Dr Unit A Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsour...

Reviews: