8. Troubleshooting
43
8.3 During a Measurement
No. Symptom
Possible Cause / Solution
Refer to
The displayed measurement value is
abnormal.
1.
Ex) Air Velocity:
Air Temp:
**.*
0.0
- If a measurement is performed out of the spec
range, “over” will be displayed on the screen.
- Probe sensor may be damaged.
2.
While measuring, the display screen
starts blinking.
- The batteries are drained.
⇒
Need new batteries.
Page 11
3.
Data cannot be stored in the internal
memory.
- Setting has not been configured properly.
⇒
MEMORY YES
Chapter
3
4.
The response of the reading is slow.
- Moving average deviations is selected.
⇒
FAST/SLOW
key
Page 9
8.4 Analog Output
No. Symptom
Possible Cause / Solution
Refer to
1. No
output.
- Output only on the initial screen or when
measuring in the measurement mode?
- The screen is on hold.
2.
The output value is wrong.
- The output range is wrong.
Page 30
8.5 Digital Output
No. Symptom
Possible Cause / Solution
Refer to
1.
DEVICE ERROR is displayed.
- The cable’s wire connection is not correct.
Page 34
2.
Data cannot be collected.
- The host computer’s setting is not properly
configured.
- Baud rate setting is not correct.
Page 32
3.
Data is wrong.
- Output format is not correct.
Page
33, 35
8.6 Printer
No. Symptom
Possible Cause / Solution
Refer to
1.
The printer turns OFF while printing.
- The printer’s NiCd battery is drained.
2.
Cannot take a hardcopy of the screen.
- The screen may not be on hold.
⇒
Put the screen on hold.
- You may be trying to take a hardcopy while
measuring.
⇒
After the calculation, try again.
Page 13
Chapter
4
3.
There is not print output after
calculation.
- Setting has not been configured properly.
⇒
PRINT YES
Chapter
4
Summary of Contents for Anemomaster 6162