3. Measurement Mode
17
START/STOP
(5)
ready
will be displayed indicating that the instrument is ready
to start a measurement.
Press
START/STOP
key to start sampling.
☆
To go back to the measurement setting screen (a), press
△
key.
[Start
Measuring]
[Waiting]
(6)
(6-1):
Sampling at one point.
(6-2):
After a sampling at one point is finished,
the instrument will standby for the next
sampling. On the screen, the average
value for one point will be displayed until
the next measurement will start.
(6-3):
One minute is left before the next
measurement starts
One minute later, the screen will
become screen (6-1) and a measurement
will start.
[Stop
Measuring]
(7)
Repeat the above procedure from (6-1) to (6-3) for the
configured number of points.
☆
To stop measuring halfway through:
START/STOP
or
▽
key.
- The data collected until the measurement is halted will be
calculated.
☆
To stop measuring (1):
△
key.
- It will go back to the screen (b) where the instrument is
ready to start a measurement.
☆
To cancel a measurement (2):
MENU
key.
- It will go back to the menu screen.
Calculation
Result
Printing
Example
"
P24
Data at each point is an average data. Average, maximum and
minimum of the average data at each point will be displayed.
☆
: hardcopy of the display screen
☆
START/STOP
or
▽
: ready for measuring (screen b)
☆
MENU
: menu will be displayed
INT (min):
Set how often a measurement to be started.
Configurable sampling interval (min):
1 ~ 6, 10, 12, 15, 20, 30, 40, 50, 60, 100, 120, 150, 180 min.
* Interval (INT) has to be set longer time than sampling time at each point.
POINTS:
Set how many points you want to perform a measurement?
Configurable number of points: 1 ~ 999 points
* For S-TIME(S), DATA(N), MEMORY, PRINT, refer to page 15 describing the average measurement.
[
Measurement Condition
]
(6-1)
<INT>
ready
T: 00/10 M:036
N: 000/010 P:016
16.8 m/s
127 ℃
<INT>
T: 01/10 M:036
N: 001/010 P:016
16.8 m/s
128 ℃
<INT>
WAIT 001 min
POINTS 001/024
16.8 m/s
126 ℃
<INT RESULT>
AVE: 19.8 m/s
MAX: 21.6 m/s
MIN : 18.5 m/s
AVE: 126 ℃
MAX: 129 ℃
MIN : 123 ℃
As for “ready” display,
refer to page 16
describing average
measurement.
(6-2)
(6-3)
(b)
WAIT 060 min
:
waiting time
remaining time length
before the next
measurement starts (min)
POINTS 001/024
: point
Current sampling
trial No.
Total set No. of
measurement trials
<INT>
WAIT 060 min
POINTS 001/024
12.8 m/s
126 ℃