5. Data Output
30
5. Data Output
5.1 Analog Output
Output on the measurement screen --------------------------------------
★
To be output every 0.25 second
★
Output Voltage: DC0 ~ 1V
★
Output Impedance: 47
Ω
5.1.1 How to Change Air Velocity Range
Display Key
Procedure
MENU
(1)
Press
MENU
key.
△
,
▽
SET
(2)
Select
5. UTILITY
.
Then press
SET
key.
△
,
▽
SET
(3)
Select
2. ANALOG OUTPUT
.
Then press
SET
key.
△
,
▽
SET
(4)
Select
1. VELOCITY
.
Then press
SET
key.
Output Range
Air Velocity
Air Temp.
0 ~ 5 m/s
0 ~ 50 °C
0 ~ 25 m/s
0 ~ 100 °C
0 ~ 50 m/s
0 ~ 200 °C
0 ~ 500 °C