![Kanomax Anemomaster 6162 Operation Manual Download Page 41](http://html.mh-extra.com/html/kanomax/anemomaster-6162/anemomaster-6162_operation-manual_652665041.webp)
5. Data Output
35
(1)
Measurement Mode (VI: Air Velocity & Air Temperature)
(2)
Calculation Mode
(AVE: Average Measurement INT: Interval Measurement FLW: Flow Measurement)
(3)
S-TIME (S) (Sampling Time)
(4)
DATA (N) (Number of Data to be collected)
(5)
POINTS (No. of samplings (partitions))
(6)
Interval Measurement: INT(measurement interval)
Flow Measurement: AREA (m2) (Duct’s cross section area)
Output Format of Measurement Condition ----- M Command
Output Format of Measurement Data ----- T Command
<Command Explanation>
Command Function
Explanation
P
To check the number of
memory pages
After the command is received, AP will be returned.
Then, the number of pages stored will be output.
N
*
To check the number of
memory data
After the command is received, AN will be returned.
Then, the number of data in the specified page (*) will be output.
M
*
To check measurement
condition
After the command is received, AM will be returned.
Then, the measurement condition in the specified page will be output.
T
*
To output measurement
data
After the command is received, AT will be returned.
Then, the measurement data in the specified page will be ouput.
*
When page number is entered incorrectly, the error message E will be returned.
As for E1 and E3, refer to
5.2.3 To Transfer Raw Data (measurement data per second)
.
* AVG Mode:
dummy output
(1)
(2)
(3)
(4)
(5)
(6)
# # ; # # # ; # # ; # # # ; # # # ; # # # # # CRLF
: 25 Characters
VT ; FLW ; 60 ; 100 ; 10 ; 1. 456CRLF
(1)
(2)
(3)
(4)
(5)
(6)
[Output
Example]
23.6; 126; *** CRLF
:
air velocity, air temp, dummy
# # # # # ; # # # # # ; # # # # # CRLF :
19 Characters
air velocity
air temp
dummy
:
(Unit: m/s,
˚
C, **)
[Output
Example]
Summary of Contents for Anemomaster 6162
Page 4: ......