![Kanomax Anemomaster 6162 Operation Manual Download Page 42](http://html.mh-extra.com/html/kanomax/anemomaster-6162/anemomaster-6162_operation-manual_652665042.webp)
6. Main Specification
36
6. Main Specification
Specification
(1)
Model Name
Main Unit:
Probe:
Model 6162
Model 0203 (for mid temperature)
Model 0204 (for high temperature)
(2)
Measurement Feature
Air velocity and air temperature (simultaneous measurement)
(3)
Measurement Object
Clean air under normal pressure and normal temperature
Air Velocity: V
0
~
50.0m/s
(4)
Measurement Range
Air Temperature:
0
~
200 °C (MODEL 0203)
0
~
500 °C (MODEL 0204)
(5)
Measurement Accuracy
(Normal Temp. Accuracy:
18~28 °C)
(6)
Temperature
Compensation Accuracy
(Air Velocity)
(7)
Heat Resistance of Cable
Teflon Coated (probe side): 200 °C
Vinyl Code (connection cable): 80 °C
(8)
Response Air
Velocity:
Air Temperature:
4 sec (90% response at air velocity of 5 m/s)
5 sec (90% response at air velocity of 5 m/s)
(9)
Display Screen
Graphic LCD (120×64dot):
displaying velocity and temperature simultaneously
w/t backlight and brightness adjustment functions
(10)
Memory Capacity
Max 999 data (when measuring in one page)
(11)
Input & Output Terminal
Remote Terminal:
Analog output Terminal:
Digital Output Terminal:
START/STOP function
simultaneous velocity & temp. output
output voltage 0
~
1V (Output Impedance: 47
Ω
)
(Output Accuracy: 0.5%FS)
RS-232C (serial)
0 ~
99 °C : V
0
= 0.2 m/s
100 ~ 199 °C : V
0
= 0.4 m/s
200 ~ 299 °C : V
0
= 0.7 m/s
300 ~ 399 °C : V
0
= 1.0 m/s
Measurement Range
Specified Accuracy Display Resolution
V
0
~ 4.99 m/s
5.00 ~ 9.99 m/s
0.01 m/s
10.0 ~ 24.9 m/s
Air
Velocity
25.0 ~ 50.0 m/s
±3%FS
0.1 m/s
0 ~ 99.9 °C
0.2 °C
100 ~ 199 °C
Air
Temp.
200 ~ 500 °C (0204only)
±1%±1
˚
C of reading
1 °C
MODEL 0203
MODEL 0204
Temp.
Velocity
0 ~ 200 °C
0 ~ 400 °C
V
0
~ 4.99 m/s
±10%FS
±15%FS
5.00 ~ 9.99 m/s
10.0 ~ 24.9 m/s
25.0 ~ 50.0 m/s
±6%FS ±10%FS
Summary of Contents for Anemomaster 6162