5. Data Output
31
△
,
▽
SET
(5)
(a)
The range highlighted is the current set range.
If there is no change to be made, press
SET
key.
The screen will go back to the ANALOG OUTPUT screen (a).
(b)
To change the range, select the range and press
SET
key.
(6)
The selected range will be highlighted, and the screen will
automatically go back to the ANALOG OUTPUT screen (a).
MENU
(7)
Press
MENU
key to go back to the menu screen.
* Analog voltage is output ony when the display screen indicates a
measurement is being performed.
To change the air temperature’s
output, perform the same
procedure.
(a)
< RANGE >
1. 0 ~ 5 m/s
2
.
0 ~ 25 m/s
3. 0 ~ 50 m/s
< MENU >
1
.MONITOR
2.MEASUREMENT
3.DATA OUTPUT
4.MEMORY CLEAR
5.UTILITY
< ANALOG OUTPUT >
1
. VELOCITY
2. TEMPERATURE
< RANGE >
1
.
0 ~ 5 m/s
2. 0 ~ 25 m/s
3. 0 ~ 50 m/s
< RANGE >
1.
0 ~ 5 m/s
2
. 0 ~ 25 m/s
3. 0 ~ 50 m/s
Range
Conversion
Equation
0~ 5 m/s
U = 0.005 × V
0~25 m/s
U = 0.025 × V
Air
Velocity
0~50 m/s
U = 0.05 × V
0~50 °C
T = 0.05 × V
0~100 °C
0~200 °C
T = 0.1 × V
T = 0.2 × V
Air
Temp.
0~500 °C
T = 0.05 × V
Velocity
Temp.
0
0
2.5
50.0
5.0 m/s
100°C
0.5
1.0
Out
p
ut
(V
)
<Example>
[Symbols]
U: Air
Velocity
(m/s)
T: Air
Temp
(°C)
V: Output
Voltage
(mV)
< Output Voltage Conversion>