Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 78 -
When the system is connected to mains supply with the power cable, the battery is charged
automatically.
A fully charged battery can work continuously for more than 1 hour. Exact time of endurance depends
on image modes and screen brightness.
Battery Specifications:
Rated voltage: 11.1V
Battery capacity: 7500mAh
Limited charging voltage: 12.6V
14.5 Data maintenance
To conduct the data management regularly, data backup to external storage medium to protect
data security. For the data, which had been backed up successfully, please timely deleted
unnecessary data within the equipment so as to avoid impacting the system performance for
long-term using.
14.6 Troubleshooting
User shall prepare a proper maintenance plan featuring regular servicing and examination in order to
ensure normal instrument performance. If any abnormity occur, timely contact International Trade Dept of
Kaixin for support.
If the following problems occur on starting up the machine, try to make corrections following the
method in the table. If the problem remains unsolved, contact International Trade Dept of Kaixin for
support.
SN
Trouble
Correction
1
No response after pressing
power on switch
1. Check power supply.
2. Check power cable and connections.
2
Character and gray scale are
displayed, but no ultrasonic
image on the screen.
Probe is not properly connected. Turn off the power and
reconnect the probe.
3
Interference like strips or
snowflake are in use display
on screen
1. Check power supply.
2. Check whether other electronic instruments
3. Check electric field or magnetic field interference in the
surrounding environment
4. Check power socket and probe whether they are connected
well
4
Not clean image in near field
Adjust total gain and TGC slider near the bottom right edge
5
Not clean image in far field
Adjust total gain and TGC slider near the bottom right edge
Warning:
Can not restore the data after delete, please cautions!
Attention:
Battery is consumable; the battery cycle-life is based on the times of charge
and discharge as unit. When the use time reduced significantly compared with normal
conditions, the battery should be promptly replaced.
Attention:
Don’t throw away the exhausted battery anywhere; especially throw it in the
fire. Please deal with it according to local statutes. Use pollution degree II to deal with.