Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 50 -
10.5 General Measurement
The measurement results of the general measurements are not recorded in the measurement report, but
all the specific measurements are based on the general measurements.
General measurements can be performed in the B mode, M mode, Color/PDI mode and PW mode. The
operations of Color/PDI mode and the B mode measurements are similar. These two modes measurements
are called BC measurement hereafter.
10.5.1 B Mode Measurements
10.5.1.1 Distance
It is used to measure the distance between two points on the image.
Steps:
1.
Press the
Caliper
key on the control panel to activate the measurement, and the measurement menu
displays on the top left of the screen.
2.
Select [Distance] on the menu by moving the trackball and press
Set
to confirm. A marker displays on
the image.
3.
Move the marker to the starting point by using the trackball, press
Set
to fix the marker and the second
marker displays.
4.
Move the second marker to the end point by using the trackball, you can also press
Change
on the
control panel to exchange the starting point and end point, press
Set
to confirm the distance between
two points, and the measurement result displays in the result box.
5.
If necessary, you can repeat step 2-4 to perform the next distance measurement.
10.5.1.2 Area and Circumference
If the target object is similar with an ellipse, ellipse measurement can be performed to measure the area
and circumference. If the target object is irregular, trace measurement can be performed to measure the
area and circumference.
Ellipse
It is used to measure the circumference and area of a blocked area on the image.
Steps:
1.
Press the
Caliper
key to activate the measurement, and the measurement menu displays on the top left
of the screen.
2.
Select [Circle/Area] on the menu by moving the trackball and press
Set
to confirm. Select[Ellipse] on
the pop-up menu. A marker displays on the image.
3.
Move the marker to the starting point by using the trackball, press
Set
to fix the marker and the second
marker displays.
4.
Move the second marker to the end point by using the trackball and press
Set
. An ellipse displays on
the image.
If you are not satisfied with the lengths of axes during measurement, press the
Change
on the control
panel to readjust.
5.
Move the trackball to adjust their lengths of axes, then press
Set
to confirm.The system automatically
calculates the area and circumference, and the measurement result displays in the result box.
6.
If necessary, you can repeat step 3-5 to perform the next area and circumference measurement.
Trace
It is used to measure the circumference and area by operating the trackball along the target object.
Steps:
1.
Press the
Caliper