Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 51 -
of the screen.
2.
Select [Circle/Area] on the menu by moving the trackball and press
Set
to confirm. Select[Trace] on
the pop-up menu. A marker displays on the image.
3.
Move the marker to the starting point by using the trackball, press
Set
to fix the marker and the second
marker displays.
4.
Move the second marker along the target object by using the trackball.
5.
Press
Set
to confirm and the system automatically calculates the area and circumference, and the
measurement result displays in the result box.
6.
If necessary, you can repeat step 3-5 to perform the next area and circumference measurement.
10.5.1.3 Volume
Three-distance
It is used to measure the volume of a cuboid shaped object by measuring the length, the width and the
depth.
Steps:
1.
Press the
Caliper
key to activate the measurement, and the measurement menu displays on the top left
of the screen.
2.
Select [Volume] on the menu by moving the trackball and press
Set
to confirm. Select [3 Dist] on the
pop-up menu. A marker displays on the image.
3.
Move the marker to the starting point by using the trackball, press
Set
to fix the marker and the second
marker displays.
4.
Move the second marker to the end point by using the trackball and press
Set
to confirm. Determine
the distance of the first line.
5.
Repeat step 3-4 to perform a distance measurement for the width.
6.
Rescan an image which is perpendicular to the previous image and repeat step 3-4 to perform a
distance measurement for the depth.The system automatically calculates the volume, and the
measurement result displays in the result box.
Ellipse distance
Steps:
1.
Press the
Caliper
key to activate the measurement, and the measurement menu displays on the top left
of the screen.
2.
Select [Volume] on the menu by moving the trackball and press
Set
to confirm. Select [EDist] on the
pop-up menu. A marker displays on the image.
3.
Move the marker to the starting point by using the trackball, press
Set
to fix the marker and the second
marker displays.
4.
Move the second marker to the end point by using the trackball and press
Set
. An ellipse displays on
the image.
If you are not satisfied with the lengths of axes during measurement, press the
Change
on the control
panel to readjust.
5.
Move the trackball to adjust their lengths of axes, then press
Set
to confirm.
6.
Rescan an image which is perpendicular to the previous image.
7.
Perform a distance measurement. The system automatically calculates the volume, and the
measurement result displays in the result box.
Ellipse
Steps:
1.
Press the
Caliper
key to activate the measurement, and the measurement menu displays on the top left
of the screen.