Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 28 -
Chapter 8 Optimizing the Image
After the patient information, probe and exam type are selected, you can select the imaging mode to
optimize the image for exact diagnosis.
8.1 B Mode
B mode is the most frequently used two-dimension imaging mode. It is intended to provide
information of anatomical structure of soft tissue and the image is in gray color in this mode.
8.1.1 Entering the B Mode
Steps for entering the B mode are as follows:
1. Select the appropriate probe and exam type, and the system automatically enters the B mode, as shown
in Figure 8-1.Or, press the
B
key on the control panel to enter the B mode from other modes.
Figure 8-1 B Mode Image
For details about the description of the main screen, refer to Section 2.6 Main Screen.
2. Optimize the image. For details, refer to Section 8.1.2 Optimizing the B Mode Image.
To save the image parameter settings as the default setting, press the
Save IP
key on the control panel.
3. Select other image modes to exit the B mode.
8.1.2 Optimizing the B Mode Image
Methods of optimizing the B mode image are as follows:
Method 1: Select the B mode parameter on the menu at the top left of the screen by using the trackball, press
the
Set
key to adjust it. Press the
Menu
key to display/hide this menu.