Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 15 -
5.4 Connect to power
There are two ways to power on the instrument: mians supply and battery.
The device's battery is automatically charged while the power is connected to the mains supply. Once
the charge is completed, the power supply will no longer continue to charge the battery.
The power supply must meet the following requirements:
100-240V~ 50Hz/60Hz
,
1.2-0.6A
Shutdown the system, insert the power plug (jack) into the power input port on the rear panel of the
main unit (Input 100-240V~ 50/60Hz), the other end to the mains socket-outlet.
The instrument uses three-core power line. It connects with the protective earth line when power plug
inserts into the power socket.
5.5 Assembling and Disassembling the battery
Disassembling and assembling the battery have a certain risk, non-technical personnel or authorized
qualified personnel by the manufacturer shall not disassemble the battery.
5.5.1 Disassembling the battery
1. Shutdown the system; hold your finger firmly on the battery cover icon and follow the arrow to
push the battery cover;
2. Remove the battery cover, use a cross screwdriver to remove the two screws on the triangular metal parts;
3. Remove the battery cable and remove the battery.
5.5.2 Assembling the battery
1. Shutdown the system; first connect the battery cable, then put the battery into the battery slot;
2. Then use a cross screwdriver to install the two screws on the triangular metal parts, and finally
install the battery cover.
5.6 Connecting the Peripherals
5.6.1 Connecting the foot switch
Shutdown the system, insert the foot-switch plug into the foot-switch socket
on the rear panel of
the main unit. Step on the foot switch to freeze/unfreeze the image.
5.6.2 Connecting the video printer
Please use the printer model recommended by the manufacturer to connect the instrument. The video
printer is purchased by the user.
1. Shutdown the system, connect the equipotential terminal ( ) of the video printer to the earthing;
Attention:
The battery life will be affected if the battery is used frequently. Please use
the mains supply as often as possible.
Warning:
1.
Do not connect the AC three-pin power cable of this instrument to a two-pin outlet
without grounding; otherwise there is a danger of electric shock.
2.
Ensure the potential-equalization lead wire is connected before inserting the equipment
power plug. Otherwise, there is danger of electric shock.
3.
The operator must not touch signal input/ signal output and patient simultaneously.
Use the foot switch provided by the manufacturer only. Otherwise, system and footswitch
may be damaged.
Attention:
The waterproof grade of foot switch is IPX1.