Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 74 -
Chapter 14 System Maintenance
The system maintenance should be performed by the user and service engineer. Users shall be in full
charge of maintenance, repair and operation of the system after purchasing the product.
14.1 Maintenance by users
14.1.1 System cleaning and disinfection
1. Clean the probe
(1) Must wear sterilized gloves to prevent possible infection.
(2) Rinse the probe with water or soapy water to remove all contaminants, or use a soft urethane
sponge to wipe the probe. Do not use brushes as it may damage the probe.
(3) After finishing the rinsing, use a sterilized cloth or gauze to wipe the water on the surface of
probe. Do not dry the probe by heating it.
2. High-level disinfection
Please follow the disinfection method provided in this user’s manual for disinfection.
(1) Before disinfection, wear sterilized gloves to prevent possible infection;
(2) You must clean the probe before disinfection. Recommend the solution to disinfect in the
following table.
Warning
: Turn off the instrument and pull out the power supply wire before cleaning
every instrument of the system. It may cause electric shock if clean the system under
power is on.
Warning
:
There is no any water-proof device in the system. Do not splash any water or
liquor into the system when cleaning or maintenance; otherwise it will cause malfunction
or electric shock.
Attention
:
1. To prevent possible infection, it is advisable to wear sterilized gloves when cleaning,
disinfecting the ultrasonic probe.
2. Clean the probe with sterile water to remove the residual chemicals after disinfection,
because the residual chemicals may be harmful for humans.
3. Kaixin Company will not make any guarantee for the efficacy of disinfector. Please
contact the appropriate manufacturer for details.
Warning
:
Do not place the ultrasonic probe connector into water or disinfectant as it
may cause electric shock or the malfunction of probe.
Attention
:
1. In the process of cleaning and disinfection, avoid probe overheat (exceeding 60
℃
) as it
may be deformed or damaged under excessive heat.
2. In the operation of disinfection, please refer to medical institutions disinfection technical
specifications.
Warning
: The following parts of the equipment are not serviced or maintained while in
use with the patient.