Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet)                                                                          User’s Manual V1.01 

 

- 47 - 

1.

 

Press the 

Body Mark

 key on the control panel to activate the body mark annotation and a body mark 

appears on the screen. 

2.

 

Select one body mark by using the trackball and press the 

Set

 key to confirm. The selected body mark 

displays on the bottom left of the screen.You can also adjust the direction of the probe by rotating the 

Angle

 knob. 

3.

 

Press 

Set

 or 

Body Mark 

key again to exit the annotation status. 

9.5.4 Deleting the Annotation 

 

Deleting the text annotation 

 

When you are adding a text annotation, move the trackball to the desired text annotation and press 

Backspace

 on the keyboard to delete one annotation. 

 

When you have added the text annotation, press 

Backspace

 to delete the annotations from the last 

one to the first one. 

Delete all the annotations:   

On all non-dialog interface, press the 

Clear

 on the keyboard to empty all annotation, including text, 

arrow, body mark, measuring scale and results.Please operate carefully! 

If you want to delete all the text and arrow annotations after freezing the image, make a setting in 

Preset screen-> [Comment Preset] -> [Clear comments while unfreezing image]. 

 

Deleting the body mark annotation 

Refer to Deleting the text annotation. 

 

Deleting the arrow annotation 

Refer to Deleting the text annotation. 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 

 

 

 

 

 

Attention:

  On  all  non-dialog  interfaces,  press  the  Clear  on  the  keyboard  to  empty  all 

annotation, including text, arrow, body mark, measuring scale and results. Please operate 
carefully!

 

Summary of Contents for DCU10

Page 1: ...KAIXIN DCU10 Full Digital Color Doppler Ultrasonic Diagnostic Instruments Vet User s Manual XUZHOU KAIXIN Electronic Instrument Co Ltd...

Page 2: ...ce in case of meeting all the following requirements 1 Assembly expansion readjustment improve and repair are all performed by professionals recognized by Kaixin 2 All replacement parts and accessorie...

Page 3: ...contains warnings for predictable dangers Users shall also exercise care at any time to be aware of the dangers unforeseen in this manual Kaixin shall not be liable for the damages and losses arising...

Page 4: ...ential danger that may result in personal injury if not avoided Attention Indicates a potential danger or unexpected use condition that may result in light injury or property loss or affecting the use...

Page 5: ...probe use Attention It is prohibited to scratch or squeeze the LCD LED Attention Do not place object on the machine Do not apply external force on the machine Close the keyboard before moving the mac...

Page 6: ...12 5 1 System installation 13 5 2 Ultrasonic probe installation 13 5 2 1 Connecting the probe 14 5 2 2 Disconnecting the probe 14 5 3 Connect to equipotential wire 14 5 4 Connect to power 15 5 5Assemb...

Page 7: ...Mode Image 36 8 4 M Mode 37 8 4 1 Entering the M Mode 37 8 4 2 Optimzing the M Mode Image 37 8 5 PW Mode 40 8 5 1 Entering the PW Mode 40 8 5 2 Optimizing the PW Mode Image 41 Chapter9 Processing the...

Page 8: ...Image Cine 70 12 2 Reviewing the Image Cine 70 12 3 Deleting the Image Cine 71 12 4 Printing the Image 71 12 5 Saving Data 71 12 5 1 Saving to USB Storage Devices 72 12 5 2 Saving to DICOM Server 72...

Page 9: ...orrelation line average space compound edge enhance image enhance acoustic power vertical reverse horizontal reverse body type 15 Color image adjustment functions gain frequency scale sensitivity base...

Page 10: ...er gallbladder pancreas spleen kidney 29 Obstetric application measurement and GA EDD calcualtion function 30 Cardiac application measurement function left ventricle inner diameter of right ventricle...

Page 11: ...ure composition operation of the keys and buttons the general work flow before using this instrument 2 1 Structure composition of the instrument DCU10 full digital color Doppler ultrasonic diagnostic...

Page 12: ...als such as monitor and projector 2 Serial port Used for connecting the debugging cable and exporting the debugging information for after sales servicing 3 Footswitch port Used for connecting the foot...

Page 13: ...charge indicator The indicator lights on when the battery is charged and lights off when the battery is fully charged Keyboard Area 1 ESC Key 1 When dialog is open press it to close the opened dialog...

Page 14: ...key Press it to adjust the volume of the frequency spectrum in the real time PW mode 17 Ctrl blank Typewriting selection key In Chinese screen press the combined keys to select the desired typewritin...

Page 15: ...djust the gain of the current mode 16 Cursor key Press it to display or hide the cursor 17 Measurement key Press it to activate or deactivate the measurement 18 Change key When measuring Press it to s...

Page 16: ...g the depth of the image 23 Magnification key Used for magnifying the interest area of the image 24 Play back key Press it to inactivate activate the manual review of the cine in the cine mode 25 Save...

Page 17: ...annotate the image Perform measurements Save the image Complete the scan and save the patient report Send data to printer server USB device End the exam Perform a scan Figure 2 6 Regular Examination...

Page 18: ...ical configuration 1 Main unit 1 unit 2 6 5 MHz micro convex probe 1 PC 3 7 5 MHz high frequency linear array probe 1 PC 4 Built in battery 1 PC 5 Power supply cable 1 PC 6 Equipotential wire 1 PC 7 N...

Page 19: ...ailure to do so may result in explosion Warning Avoid using this equipment with high frequency electric knife high frequency therapy equipment or defibrillators and other electronic devices or may an...

Page 20: ...ilure the patient will have danger of electric shock 6 Additional equipment connected to the medical electrical equipment must comply with the respective IEC or ISO standards e g IEC60950 for data pro...

Page 21: ...ck the ultrasonic probe and connecting cable after diagnostic operation Use of defective probe may cause electric shock 3 Do not strike the probe using the damaged probe may cause electric shock to th...

Page 22: ...ing the probe Disconnecting the probe with system power on may damage the system or probe 2 Before disconnecting the probe place the probe on the probe holder and put its cable on the probe cable hang...

Page 23: ...ery cable and remove the battery 5 5 2 Assembling the battery 1 Shutdown the system first connect the battery cable then put the battery into the battery slot 2 Then use a cross screwdriver to install...

Page 24: ...en the user does not perform any operaton within the set time Go to sleep Press the Power on Standby key in the upper left corner of the control panel to exit the standby mode Powering off the system...

Page 25: ...tions 7 See to it that there are no foreign objects on the control panel Remove the object if any 8 No barriers around the intake of equipment 9 See to it that probe has been cleaned disinfected else...

Page 26: ...one of the menu items and press the Set key to enter the screen of this menu item Move the trackball to at the right of one menu item and press the Set key the submenu items are displayed Move the tra...

Page 27: ...image Options JPG BMP or FRM Format Film Type Set the saving format of cine Options CIN orAVI Format TI Set the TI index Options TIS TIB or TIC Video Size Set the resolution of the printer Options 64...

Page 28: ...ode and set the time to enable the standby mode Probe Set the default probe to be used when the system is powered on Reload Preset on New Patient Enable or disable the function of retrieving the saved...

Page 29: ...one template 2 Input the defined name for the selected exam template at Exam Mode and click Add to add it The defined exam type displays in the list below Select exam type in the list and click Renam...

Page 30: ...22 Figure 6 5 Measurement Preset Screen Item Descriptions HR Set the heart rate Options 1 8 Cursor Line Display Enable or disable the dotted line between two markers 6 5 Presetting the Annotations You...

Page 31: ...f custom and click Move Up or Move Down to move it up or down To delete an annotation Select one annotation in the list of The comment of custom and click Delete to delete it To select the language of...

Page 32: ...e subnet mask of the system Gateway Enter the gateway of the system Server AE title Enter the name ofAE title Host Enter the host name of the server IP address Enter the IP address of the server Port...

Page 33: ...report template to be deleted click Delete pop up warning box click OK to delete the selected report template Click Delete all pop up warning box click OK to delete all the report templates To import...

Page 34: ...ame owner name patient ID gender and date of birth by the keyboard 2 When measuring cardiac you can input the patient s height and weight BSAwill be calculated by the system automatically You also can...

Page 35: ...am and the system enters into Patient Info screen to start the next exam Or press the Patient key select New Patient and OK to start the next exam Attention Probe types and icons display on the LED sc...

Page 36: ...type and the system automatically enters the B mode as shown in Figure 8 1 Or press the B key on the control panel to enter the B mode from other modes Figure 8 1 B Mode Image For details about the de...

Page 37: ...d to adjust the gain that allows compensation for attenuation of the echoes over time depth It is adjustable during the real time scan regardless of the imaging mode and display format TGC balances th...

Page 38: ...enu item press Select knob item to confirmation select Colorize item and then rotate Select knob to choose the color 8 1 2 8 Adjusting the Dynamic Range Dynamic range increases the adjustable range of...

Page 39: ...the LineAverage Line average is used to adjust smoothness of the image A higher line average brings lower contrast and smoother image Steps Move cursor to Line Average menu item and press Set key to a...

Page 40: ...it 8 1 2 19Adjusting the Horizontal Reverse Horizontal reverse is used to adjust two dimensional image flip left right Steps Move cursor to Horizontal Reverse menu item and press Set key to adjust it...

Page 41: ...t Press the Menu key to display hide this menu Method 2 Select the Color mode parameter on the menu at the top left of the screen by using the Select knob press Select knob change state between parame...

Page 42: ...eed The flow speed that can be detected in the Color mode is adjusted Steps Move cursor to Flow Speed menu item and press Set key to adjust it Rotate Select knob to highlight the Flow Speed menu item...

Page 43: ...present the different flow direction respectively Steps Move cursor to Color Map menu item and press Set key to change the color Rotate Select knob to highlight the Color Map menu item press Select kn...

Page 44: ...t the size of the ROI or press the Set key again to adjust the position of ROI 4 Optimize the PDI mode image For details refer to Section 8 3 2 Optimizing the PDI Mode Image Press the Save IP key to s...

Page 45: ...rackball to adjust the position of the M line 4 Optimize the M mode image For details refer to Section 8 4 2 Optimzing the M Mode Image Press the Save IP key to save the parameter settings in the M mo...

Page 46: ...nd then rotate Select knob to adjust it 8 4 2 4 Adjusting the Focus Position Focal position is used to adjust the position of the ultrasound beam The focus displays as a triangle on the scale Steps Mo...

Page 47: ...to confirmation select Line Average item and then rotate Select knob to adjust it 8 4 2 10 Adjusting the Edge Enhance Edge enhance is used to distinguish the image outline and optimize the smoothness...

Page 48: ...measuring the velocity X axis Indicates the time Y axis Indicates Doppler frequency scale including a positive and negative indicator 3 Adjust the position and angle of the spectral Doppler line Move...

Page 49: ...knob change parameter item for parameter item select state or paramter value for parameter item adjust state Press the Menu key to display hide this menu Parameters Adjustment in the PW Mode 8 5 2 1 A...

Page 50: ...Angle item and then rotate Select knob to adjust it 8 5 2 7Adjusting the Scan Speed Time scale and the refresh speed of the spectrum vary with the scan speed Steps Move cursor to Sweep Speed menu item...

Page 51: ...stable range of contrast by enhancing the intensities of parts of the grayscale Steps Move cursor to Dynamic Ra menu item and press Set key to adjust it Rotate Select knob to highlight the Dynamic Ra...

Page 52: ...play The image is spitted by two parts with a real time scan on the left and no image on the right 3 Press again to freeze the real time scan and unfreeze the frozen image The probe direction mark is...

Page 53: ...t the position of ROI by moving the trackball If you are not satisfied with the size press the Set key again to adjust the size of ROI 5 Press the Zoom key to magnify the interest region You can adjus...

Page 54: ...he frozen or real time scan 9 5 1 Text Annotation Steps are as follows 1 Press the Comment key on the control panel to activate the annotation Annotation library appears on the left of the screen 2 Se...

Page 55: ...Backspace on the keyboard to delete one annotation When you have added the text annotation press Backspace to delete the annotations from the last one to the first one Delete all the annotations On a...

Page 56: ...2 Measurement Menu Measurement Area The user performs a measurement in the ultrasound image area Measurement Result box After the user makes a measurement the measurement and calculation results are...

Page 57: ...ng sections are described according to the measurement menu 10 3 Measurement Preset You can preset the measurement For details refer to Section 6 4 Presetting the Measurements 10 4 Measurement Error U...

Page 58: ...the next distance measurement 10 5 1 2Area and Circumference If the target object is similar with an ellipse ellipse measurement can be performed to measure the area and circumference If the target ob...

Page 59: ...end point by using the trackball and press Set to confirm Determine the distance of the first line 5 Repeat step 3 4 to perform a distance measurement for the width 6 Rescan an image which is perpendi...

Page 60: ...to fix the marker and the second marker displays 4 Move the second marker to the end point by using the trackball and press Set An ellipse displays on the image If you are not satisfied with the lengt...

Page 61: ...second marker to the end point by using the trackball and press Set Determine one line of the included angle 5 Repeat step 3 4 to draw a second line to create an angle The system automatically calcul...

Page 62: ...Set to confirm A marker displays on the image 4 Move the marker to the starting point by using the trackball press Set to fix the marker and the second marker displays You can also press Change on th...

Page 63: ...oppler waveform Steps 1 Press the PW key on the control panel to enter the PW mode 2 Press the Caliper key to activate the measurement and the measurement menu displays on the top left of the screen 3...

Page 64: ...clinical diagnosis purposes by tracing one or more Doppler waveforms Manual trace and auto trace are available Manual trace Steps 1 Press the PW key on the control panel enter the PW mode then press F...

Page 65: ...e patient information is input Confirm that the exam type is selected 10 6 1 Abdomen Measurements Abdominal measurements can be performed in B Color PDI mode Steps 1 Press the Measure key in the real...

Page 66: ...rm 3 Follow the methods in the following table to perform the measurement and the measurement results display in the result box Type Measurement Item Description Measurement Method Cat BD Body Diamete...

Page 67: ...erformed in the B Color PDI mode or M mode 10 6 3 1 Cardiology Measurements in the B Color PDI Mode Cardiology measurements in the B Color PDI mode include left ventricle measurement left ventricle we...

Page 68: ...mages covering the long axis and the short axis of the left ventricle Steps 1 Press the Measure key in the real time or frozen B Color PDI mode and the measurement menu displays on the top left of the...

Page 69: ...on the image 4 Perform the LVLd LVLs LVAMd LVAMs measurements one by one and the measurement results displays in the result box Measurement Item Measurement Sub item Description Measurement Method Lef...

Page 70: ...at end systole LVAMd LV short axis area at mitral value at end diastole Refer to Section 10 5 1 2 Area and Circumference LVAMs LV short axis area at mitral value at end systole LVAPd Area of LV short...

Page 71: ...nu displays on the top left of the screen 2 Select RV on the menu by moving the trackball and press Set to confirm A marker displays on the image 3 Perform a distance measurements and the measurement...

Page 72: ...e measurement aorta measurement ejection time measurement LVMW measurement and HR measurement 10 6 3 9 Left Ventricle The methods for measuring left ventricle in the M mode are as follows Teichholz me...

Page 73: ...ng the trackball and press Set to confirm 3 Select Cube on the pop up menu and a marker displays on the image 4 Perform the LVIDd LVIDs measurements one by one and the measurement results displays in...

Page 74: ...tomatically calculates the mitral valve volume by the following formula and the measurement results displays in the result box QMV ml 4 DEV cm s DCT s Measurement Item Description Measurement Method D...

Page 75: ...on the image 3 Perform the LVPWd IVSTd LVIDd measurements one by one and the measurement results displays in the result box Measurement Item Measurement Sub item Description Measurement Method Left Ve...

Page 76: ...al time or frozen B Color PDI mode and the measurement menu displays on the top left of the screen 2 Select Lt Thyroid or Rt Thyroid on the menu by moving the trackball and press Set to confirm 3 Perf...

Page 77: ...report preset see 6 7 Preset of Report Template 11 2 2 Adding Ultrasound Images Image s saved in current exam can be added to the report 1 In the report screen click the Add Image to pop up the add i...

Page 78: ...all and press the Set key to save the cine 12 2 Reviewing the Image Cine Current patient You can view the saved images or cines in the Review screen Steps are as follows 1 Press the Review key on the...

Page 79: ...the image or cine 4 Press the Freeze key to return to the main screen 12 3 Deleting the Image Cine You can delete the saved image or cines in the Review screen 1 Press the Review key on the control p...

Page 80: ...COM Server Current patient images Steps are as follows 1 Press Review to enter the Review screen and select the images Or click Select All or enable Mult Sele at the left to select the images 2 Click...

Page 81: ...L 3 Move the cursor to Visiable press the Set key to turn on or turn off the puncture guide line Move the cursor to Position press the Set key to set the position of the puncture guide line BGL 1 BGL...

Page 82: ...the instrument and pull out the power supply wire before cleaning every instrument of the system It may cause electric shock if clean the system under power is on Warning There is no any water proof d...

Page 83: ...erile water about 2 gallons for at least one minute to remove the residual chemicals You may follow the recommended method by the disinfectant manufacturer to rinse 4 After finishing the rinsing use a...

Page 84: ...placing the power supply cord please contact us replace the power supply cord under the guidance of our company Please use the power supply cord provided by us 14 4 Using the Battery 14 4 1 Precaution...

Page 85: ...he leakage explosion or fire 13 Do not use the abnormal battery with particular smell or abnormal heat or distortion or turn colors or abnormal phenomena which will lead to the leakage explosion or fi...

Page 86: ...e If the problem remains unsolved contact International Trade Dept of Kaixin for support SN Trouble Correction 1 No response after pressing power on switch 1 Check power supply 2 Check power cable and...

Page 87: ...please consult the International Trade Dept of Kaixin or its dealers as they need to have professional technology engineers The following safety checks should be performed at least every 12 months by...

Page 88: ...nd snow nor shall the goods be placed upside down bumped knocked or over stacked Chapter 16 Safety Classification 1 Classified according to the type of protection against electric shock Class I EQUIPM...

Page 89: ...n Mansion C 01 Economic Development Zone Xuzhou Jiangsu China Post Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website http www kxele com E mail info kxele com Information co...