8 Startup and parameterization
32
8.2
Operating levels
The device's operating levels consist of a process level and configuration level.
Level
Functions
Process level
Extracting the flow rate measured by the device, the current value
emitted at the 4 to 20 mA current output, and the values for the
main counter and daily counter
Extracting the daily counter
Switching to the configuration level
Configuration level
with parameterization, testing, and
information menu
Adjusting the device's parameters
Testing certain device parameters
Calibrating the device
Extracting any warnings or error messages generated by the de-
vice if the device status LED is orange or red
Function
Basic setting
LANGUAGE
English
UNIT of the flow rate
l/min.
UNIT for the counter
Liters
K-FACTOR
1000
OUTPUT AO1
4 mA = 0.000
20 mA = 0.000
OUTPUT DO1
Pulse
PU = 0.00 liters
OUTPUT DO2
Hysteresis
2- = 0.000
2+ = 0.000
Not inverted
Delay = 0
OUTPUT DO3
Hysteresis
3- = 0.000
3+ = 0.000
Not inverted
Delay = 0
INPUT DI1
Inactive
FILTER
5, slow
FREQUENC.
50 Hz
CUT-OFF
0.000
BACKLIT
Filter 9, activation period: 30 s
K-SENSOR
KW = 1.000
FLOW-W.
W- = 0.000
W+ = 0.000