65
8 Startup and parameterization
8.7
Test menu details
To access the test menu, press the
+
+
keys at the same time for more than 5 seconds.
This menu enables:
CAL. AO1
The current output (4 to 20 mA) to be adjusted.
CALIB.0
The flow zero point of the device to be calibrated.
FLOW
The outputs to be checked for proper operation.
K-SENSOR
The KW value of the sensor to be changed in order to adjust the device's
accuracy.
FLOW-W.
The flow rate range to be adjusted outside which a warning message is
generated.
END
SAVE N/Y
12.6 L/S
Saving the changes made within the test
menu or not. If the changes are saved, the de-
vice operates with the new settings.
Process level