8 Startup and parameterization
Calibrating the flow rate zero point using the CALIB 0 function in the test menu
1. If you do not wish to change any further parameters, go to the END function in the test menu and
press the
key in order to save the settings and return to the process level.
1. Charge the pipe.
2. Stop medium circulation.
CALIB.0
CALIB.N
CALIB.Y
1. Choose CALIB.Y.
MEASURE /
The device calibrates the flow rate zero point.
FLOW
CAL.FAIL
The calibration of
the flow rate zero
point has succeed-
ed.
The device has detected
a non-nil flow rate in the
pipe.
The device status LED
turns orange.