Appendix
96
Device modes
Parameter
Setting
out_mode_01
x
Setting 0 (for reasons of downwards compatibility
the parameters 1 or 2 can also be used. Within the
device it has no effect)
out_mode_02
x
Self-tuning:
0 = Self-tuning disabled
2 = Self-tuning enabled
out_mode_03
x
Setting of the operating mode EPROG:
0 = 0-10 V
1 = 0-20 mA
out_mode_04
x
Setting of the temperature control:
0 = Internal control
1 = External control
out_mode_05
x
Start/stop command of the device in remote control
mode:
0 = Stop tempering
1 = Start tempering
out_mode_08
x
Setting of the internal control dynamic:
0 = non-periodic
1 = Standard
out_mode_11
x
Actuating variable source:
0 = Thermostat
1 = Serial
2 = Analog (EPROG)
System parameter Parameter
Setting
out_par_04
xxx
Setting the Cospeed control parameter for internal
control
out_par_06
xxx
Setting of the control parameter XP for internal
control
out_par_07
xxx
Setting of the control parameter TN for internal
control
out_par_08
xxx
Setting of the control parameter TV for internal
control
out_par_09
xxx
Setting of the control parameter XP of the
subordinate controller (cascade temperature control)
out_par_10
xxx
Setting of the control parameter XPU of the
subordinate controller (cascade temperature control)