12.3.4 Holding registers
Register
address
Protocol
address
Data type Explanation
Adjustable range
40001
0
ushort
Start/stop the device
0: Device is in standby
1: Device is started
40002
1
ushort
Type of actuating variable
0: Controller
1: Digital
2: EPROG
40003-
40004
2-3
float
Setpoint temperature
min. - max.
40005
4
short
Actuating variable via
Modbus
-100 … 100
Control parameter
40011
10
ushort
Internal/external
temperature control
0: Temperature control
in the internal bath
1: Temperature control
with external Pt100
40012
11
ushort
Set identification
0: No identification.
The saved parameters
are used for the
regulation
1: Unique
identification of the
controlled process
after the next start
2: Permanent
identification of the
controlled process
after each start
40013-
40014
12-13
float
Xp control parameter of the
internal controller.
0.1 … 99.9 K
40015
14
ushort
Tn control parameter of the
internal controller.
3 … 9999 s
40016
15
ushort
Tv control parameter of the
internal controller.
0 … 999 s
40017
16
ushort
Control dynamics
0: Aperiodic
1: Standard
40018-
40019
17-18
float
Xp control parameter of the
cascade-controller.
0.1 … 99.9 K